Measurement and Evaluation of the Safety of UV Light Sources|Part5: UV Lamp for Health Protection

UV lamp for health protection is used in speckle reducing and hairdressing and so on. We put the UV lamp for health protection in the UV radiation box to test the UV spectral irradiance distribution.

Test condition: the distance from the light receiving head to the light source center is 1m. The tube current is 0.0823A, the tube voltage is 219.8V, and the power is 18.1W.
Test result: spectral radiation distribution is as shown in figure 5
Illuminance: 6.25 lx;
Radiation intensity of illuminance: 11.4uW/cm2
Photochemical harm effective radiation intensity of illumination of skin and eyes: 0.5309uW/cm2
Near UV harm effective radiation intensity of illumination of eyes: 6.5284uW/cm2
250~320nm B band effective radiation intensity of illumination (E1):1.0682172uW/cm2320~400nm A band effective radiation intensity of illumination (E2):0.005564uW/cm2

a) Photochemical harm of skin and eye

Photochemical harm of skin and eye

b) Near UV harm of eye

Near UV harm of eye

From formula(2a) (2b), we can get:

Spectral distribution of UV lamp for health protection

Figure 5: Spectral distribution of UV lamp for health protection

Safety evaluation curve of UV lamp for health protection

Figure 6: Safety evaluation curve of UV lamp for health protection

Similarly, we can draw up the safety evaluation curve based on the above formula, as shown in figure 6:

a) In order to avoid the harm of the two kinds of UV light under the UV lamp for health protection, the lamp should work in the public areas below the two curves.

b) Example: when human body irradiation time is 1000s, the light illumination of the human body illuminating area should be controlled less than 35.317lx. If we use the lamp under the illuminance of 10lx, the time for people receiving the illumination is less than 353.2s.

c) We can get the following conclusion according to the requirement of using UV instruments by IEC standard:

UV instruments type

Effective radiation illumination(W/m2)

Corresponding luminance (lx)

Safety irradiation time (s)

UV type 3

E1<0.15, E2<0.15

E<87.76

t is confirmed by (5)

UV type 4

E1≥0.15, E2<0.15

E≥87.76

t≤402.4 is resulted from (5)

About Lisun Group:

Lead in CFL and LED Tester. Click to our product lists:

Goniophotometer
Spectroradiometer
Integrating Sphere
Colorimeter and Photometer
LED Test Instruments
CFL Testing Instruments
EMC testing
Electronic Ballast Tester
Equipments for Testing Electronic components
Electrical Safety Tester
Environmental Chamber
AC and DC power supply
Spectrophotometer

Leave Message

Your email address will not be published. Required fields are marked *

=